UNIT 2: BIOMEDICAL ELECTRONICS – COMPREHENSIVE SHORT NOTES
1. FUNDAMENTALS OF HUMAN PHYSIOLOGY AND BIOELECTRICITY
1.1 Cell Structure and Function
-
Cell Membrane: Phospholipid bilayer with embedded proteins; selectively permeable; maintains resting membrane potential.
-
Organelles: Nucleus (DNA), mitochondria (ATP), endoplasmic reticulum (protein/lipid synthesis).
-
Resting Membrane Potential (RMP): Typically -70 mV (inside negative). Established by Na⁺/K⁺ ATPase pump (3 Na⁺ out, 2 K⁺ in) and leak channels.
1.2 Bioelectric Potentials
- Resting Potential (Nernst Equation):
$$E_{ion} = \frac{RT}{zF} \ln \left( \frac{[ion]_{out}}{[ion]_{in}} \right)$$
At 37°C, simplifies to: $$\displaystyle E_{ion} = \frac{61.5}{z} \log \left( \frac{[ion]_{out}}{[ion]_{in}} \right) $$ (mV).
[!TIP] For K⁺ (z=1), typical [K⁺]ₒ/[K⁺]ᵢ ≈ 1/20 → Eₖ ≈ -90 mV.
-
Action Potential (AP):
-
Phases:
-
Depolarization: Na⁺ channels open → rapid rise to +30 mV.
-
Initial Repolarization: Na⁺ channels inactivate, K⁺ channels open.
-
Plateau (cardiac only): Ca²⁺ channels open, balance K⁺ efflux.
-
Repolarization: K⁺ efflux dominates → return to RMP.
-
Hyperpolarization: K⁺ channels slow to close → membrane potential briefly more negative.
-
-
Propagation: Local current flow depolarizes adjacent membrane.
-
Refractory Periods:
-
Absolute: No stimulus can trigger AP (Na⁺ channels inactivated).
-
Relative: Strong stimulus can trigger AP (some Na⁺ channels recovered).
-
-
1.3 Major Body Systems (Functional Relevance)
-
Cardiovascular: Heart generates ECG; blood pressure, flow, volume measurements.
-
Respiratory: Lung volume/flow (spirometry), gas exchange (capnography, blood gases).
-
Nervous: Brain generates EEG, EMG (muscle), nerve conduction studies.
2. BIOELECTRIC SIGNALS: CHARACTERISTICS AND CLINICAL SIGNIFICANCE
2.1 Electrocardiogram (ECG)
-
Waveform Components:
| Component | Origin | Duration (s) | |-----------|--------|--------------| | P wave | Atrial depolarization | 0.08–0.11 | | PR interval | AV nodal delay | 0.12–0.20 | | QRS complex | Ventricular depolarization | 0.06–0.10 | | ST segment | Ventricular plateau | 0.08–0.12 | | T wave | Ventricular repolarization | 0.16–0.24 | | QT interval | Total ventricular activity | <0.44 (rate-corrected) | | U wave | Possible Purkinje repolarization | 0.02–0.04 |
-
Lead Configurations:
-
Limb Leads (I, II, III, aVR, aVL, aVF): Einthoven's triangle.
-
Precordial (V1–V6): Horizontal plane.
-
-
Clinical Interpretation:
-
Arrhythmias: Tachycardia/bradycardia, atrial fibrillation (no P waves, irregular R-R).
-
Ischemia/Infarction: ST elevation/depression, Q waves, T wave inversion.
-
2.2 Electroencephalogram (EEG)
-
Frequency Bands:
| Band | Frequency (Hz) | State/Condition | |------|----------------|-----------------| | Delta | <4 | Deep sleep, coma, brain injury | | Theta | 4–8 | Drowsiness, meditation, some seizures | | Alpha | 8–13 | Relaxed, eyes closed (occipital) | | Beta | 13–30 | Alert, anxious, active thinking | | Gamma | >30 | Cognitive processing, perception |
-
Electrode Placement: 10–20 system (based on skull measurements: nasion-inion, preauricular points).
[!TIP] Fp = frontal pole, C = central, P = parietal, O = occipital, T = temporal, Z = midline.
2.3 Other Bioelectric Signals
-
EMG (Electromyogram): Muscle electrical activity; needle/surface electrodes; motor unit action potentials.
-
EOG (Electrooculogram): Corneal-retinal dipole; eye movement/blink detection.
-
ERG (Electroretinogram): Retinal response to light; photopic (cones), scotopic (rods).
3. ELECTRODES AND TRANSDUCERS
3.1 Electrode Theory
-
Half-Cell Potential: Electrochemical potential at metal-electrolyte interface; causes DC offset.
-
Polarization: Build-up of reaction products at electrode surface; increases impedance over time.
-
Impedance: $$\displaystyle Z = R + jX $$; low and stable impedance crucial for signal quality.
-
Skin Preparation: Abrasion, cleaning, electrode gel to reduce impedance and motion artifact.
-
Electrode Paste/Gel: Conductive (Ag/AgCl), high water content, low polarization, neutral pH.
3.2 Types of Electrodes & Selection
| Type | Description | Applications |
|---|---|---|
| Surface | Metal plate with gel; non-invasive | ECG, EEG, EMG |
| Needle | Insulated wire, exposed tip; invasive | EMG, nerve conduction |
| Micro | Very small (µm); for single cells | Research |
| Array | Multiple electrodes in grid | EEG, ECoG |
| Floating | High input impedance buffer at electrode | Reduces motion artifact |
Selection Criteria: Invasiveness, frequency response, impedance, comfort, longevity.
3.3 Transducers
-
Definition: Device converting one energy form to another (e.g., physiological → electrical).
-
Classification:
-
Active (generate own signal): Piezoelectric, thermocouple.
-
Passive (require external power): Strain gauge, capacitive.
-
Primary (direct): Thermocouple.
-
Secondary (indirect): Strain gauge (measures strain → force).
-
-
Piezoelectric: Certain crystals (PZT) generate charge under stress; used in ultrasound transducers, pressure sensors.
-
Temperature:
-
Thermocouple: Two dissimilar metals → Seebeck voltage (mV/°C).
-
RTD: Resistance increases with temperature (Pt100: 100 Ω at 0°C).
-
Thermistor: Semiconductor; NTC (resistance ↓ with T) or PTC.
-
-
Optical:
- Photodiode/Phototransistor: Light → current; used in pulse oximetry (photoplethysmography).
-
Resistive/Capacitive:
-
Strain gauge (resistive): ΔR/R ∝ strain.
-
Capacitive pressure sensor: ΔC ∝ 1/distance.
-
4. SIGNAL CONDITIONING FOR BIOMEDICAL SIGNALS
4.1 Biomedical Amplifiers
-
Requirements:
-
High Input Impedance (>100 MΩ): Prevents loading high-impedance sources (e.g., electrodes).
-
Low Noise: <1 µV rms input-referred.
-
High CMRR (>80–100 dB): Rejects common-mode (e.g., 50/60 Hz power line).
-
Safety Isolation: Patient protection (IEC 60601).
-
-
Instrumentation Amplifier (Three-Op-Amp Topology):
-
First stage: Two buffers → high input impedance.
-
Difference amplifier: Rejects common-mode, amplifies differential.
-
Gain set by single resistor: $$\displaystyle G = 1 + \frac{50\text{k}\Omega}{R_G} $$ (typical AD620).
-
-
ECG Front-End: Typically 3-lead or 5-lead; bandwidth 0.05–150 Hz; high CMRR for limb leads.
4.2 Filters
-
Low-Pass: Anti-aliasing (cutoff <½ sampling rate); e.g., 150 Hz for ECG.
-
High-Pass: Remove baseline wander (e.g., 0.05 Hz for ECG).
-
Band-Pass: Combine LPF and HPF; e.g., EEG: 0.5–70 Hz.
-
Notch: 50/60 Hz power line interference; active twin-T or digital.
-
Active vs Passive: Active (op-amp) provides gain and high input impedance; passive (RC) attenuates.
4.3 Isolation and Safety
-
Optical Isolators: LED + phototransistor; electrical isolation via light.
-
Transformer Isolation: AC-coupled; barrier for patient leakage.
-
IEC 60601 Leakage Current Limits:
| Type | Application | Max Leakage (µA) | |------|-------------|------------------| | B | Non-patient contact | 100 | | BF | Patient contact (except heart) | 100 | | CF | Direct cardiac connection | 10 |
5. CARDIOVASCULAR SYSTEM MEASUREMENTS
5.1 Blood Pressure Measurement
-
Non-Invasive:
-
Auscultatory: Korotkoff sounds (Phase I = systolic, Phase V = diastolic); mercury/aneurid sphygmomanometer.
-
Oscillometric: Cuff pressure oscillations; algorithm detects mean pressure, then systolic/diastolic.
-
-
Invasive: Catheter-tip transducer (strain gauge or capacitive); dynamic response must be critically damped to avoid overshoot.
5.2 Heart Rate and Rhythm Monitoring
-
From ECG: QRS detection (amplitude/threshold); RR interval → heart rate.
-
From Pulse (PPG): Photoplethysmography; peak detection.
-
Heart Rate Variability (HRV): Analysis of RR interval variations; time-domain (SDNN, RMSSD) and frequency-domain (LF, HF) metrics; autonomic nervous system assessment.
5.3 Heart Sounds (Phonocardiography)
-
S1: AV valve closure (lub); start of systole.
-
S2: Semilunar valve closure (dub); start of diastole.
-
S3: Ventricular filling (pathological in adults).
-
S4: Atrial kick (pathological).
-
Murmurs: Turbulent flow; systolic/diastolic timing indicates valve disorder.
-
Electronic Stethoscopes: Amplify, filter (20–2000 Hz), may include digital processing for noise reduction.
5.4 Photoplethysmography (PPG) and Pulse Oximetry
-
Principle: Light (660 nm red, 940 nm IR) transmitted/reflected through tissue; blood absorption varies with pulse volume.
-
Pulse Oximeter:
- SpO₂ Calculation: Beer-Lambert Law: $$\displaystyle I = I_0 e^{-\epsilon c d} $$; ratio of ratios:
$$\text{SpO}_2 = \frac{\text{AC}_{\text{red}}/\text{DC}_{\text{red}}}{\text{AC}_{\text{IR}}/\text{DC}_{\text{IR}}}$$
Calibrated against arterial blood gases.
- Limitations: Motion artifact, low perfusion, nail polish, dyshemoglobins (COHb, MetHb).
6. RESPIRATORY SYSTEM MEASUREMENTS
6.1 Spirometry
-
Parameters:
-
Vital Capacity (VC): Max volume exhaled after max inhalation.
-
Forced Vital Capacity (FVC): VC exhaled forcefully.
-
Forced Expiratory Volume in 1 s (FEV₁): Volume exhaled in first second of FVC.
-
FEV₁/FVC Ratio: <70% indicates obstructive disease (e.g., asthma).
-
-
Types:
-
Volume-Measuring: Water bellows or dry rolling seal; direct volume.
-
Flow-Measuring (Pneumotachograph): Measures flow (ΔP across resistor); volume = ∫ flow dt.
-
6.2 Photo Spirometer
-
Working: LED/photodiode pair detects vane position in airflow; vane angle ∝ flow; integrated for volume.
-
Advantages: No moving seals, compact.
6.3 Additional Measurements
-
Capnography: End-tidal CO₂ (EtCO₂) via infrared absorption; monitors ventilation.
-
Respiratory Rate: From flow/pressure waveform or impedance pneumography.
7. BLOOD GAS AND CHEMICAL ANALYSIS
7.1 Partial Pressure of Oxygen (PO₂)
-
Clark Electrode (Polarographic):
-
Electrochemical Cell: Pt cathode, Ag/AgCl anode, KCl electrolyte, O₂-permeable membrane.
-
Reaction: O₂ + 2H₂O + 4e⁻ → 4OH⁻; current ∝ PO₂.
-
Calibration: Two-point (0% and 100% O₂); temperature compensation required.
-
Drift: Membrane degradation, electrolyte depletion.
-
7.2 pH, PCO₂, Electrolytes
-
Ion-Selective Electrodes (ISE):
-
pH: Glass membrane (H⁺-selective); potential vs. reference follows Nernst.
-
PCO₂: Severinghaus electrode (pH-sensitive ISE behind CO₂-permeable membrane; CO₂ hydration changes pH).
-
Na⁺, K⁺, Ca²⁺: Liquid membrane or solid-state ISEs.
-
-
Blood Gas Analyzers: Combine Clark, Severinghaus, ISEs; sample handling (heparinized syringe, temperature control).
8. MEDICAL IMAGING MODALITIES
8.1 Magnetic Resonance Imaging (MRI)
-
Principles:
-
Nuclear Magnetic Resonance: Protons align with B₀ field; RF pulse at Larmor frequency $$\displaystyle f_0 = \frac{\gamma B_0}{2\pi} $$ (γ = 42.58 MHz/T for H).
-
Relaxation:
-
T1 (spin-lattice): Longitudinal recovery; fat = short T1 (bright on T1-weighted).
-
T2 (spin-spin): Transverse decay; fluid = long T2 (bright on T2-weighted).
-
-
-
System Components:
-
Main Magnet: Superconducting (liquid He), 1.5–3 T typical.
-
Gradient Coils: Spatial encoding (slice selection, frequency/phase encoding).
-
RF Transmitter/Receiver: Excite spins, receive signal.
-
-
Safety:
-
Ferromagnetic objects: Projectile hazard.
-
Specific Absorption Rate (SAR): RF energy deposition; limits to prevent heating.
-
Acoustic Noise: Gradient coil vibration; hearing protection required.
-
8.2 Endoscopy
-
Types:
-
Rigid: Metal tubes; laparoscopy, arthroscopy.
-
Flexible:
-
Fiberscope: Image via fiber bundle; lower resolution.
-
Video endoscope: CCD/CMOS at tip; digital signal transmission.
-
-
-
Image Transmission: Fiber optics (coherent bundle) or electrical (video).
-
Applications: GI tract (esophagoscopy, colonoscopy), bronchoscopy, cystoscopy.
8.3 Thermography
-
Infrared Imaging: Detects radiant heat (8–14 µm wavelength).
-
Applications: Inflammation (arthritis), vascular disorders (deep vein thrombosis), breast cancer screening (as adjunct).
9. THERAPEUTIC EQUIPMENT
9.1 Defibrillators
-
Types:
-
DC: Capacitor discharge (monophasic/biphasic); most common.
-
AC: Outdated (60 Hz sine wave).
-
AED: Automated external; voice prompts, automatic rhythm analysis.
-
-
Synchronization: Synchronized shock (R wave detection) for cardioversion; unsynchronized for VF/VT.
-
Waveforms: Biphasic (more effective, less myocardial damage) vs. monophasic.
-
Energy Dosing: Biphasic: 120–200 J; Monophasic: 360 J. Pediatric: 2–4 J/kg.
9.2 Pacemakers
-
Types:
-
Fixed-rate: Constant rate (VOO, AOO).
-
On-demand:
-
VVI: Ventricular pacing, ventricular sensing, inhibition.
-
DDD: Dual-chamber pacing and sensing; maintains AV synchrony.
-
Biventricular (CRT): For heart failure; LV and RV pacing.
-
-
-
Components: Pulse generator (battery + circuitry), leads (unipolar/bipolar), programmer.
-
Modes: NBG code ( chambers paced, sensed, response).
9.3 Heart-Lung Machine (Cardiopulmonary Bypass)
-
Components:
-
Blood Pump: Roller or centrifugal; non-pulsatile or pulsatile flow.
-
Oxygenator: Bubble (direct gas contact) or membrane (semipermeable; gas exchange across membrane).
-
Heat Exchanger: Controls blood temperature (hypothermia).
-
Filters: Remove air, debris, clots.
-
-
Operation: Venous blood drained → oxygenated → temperature controlled → returned to arterial system; heart arrested (cardioplegia).
9.4 Hemodialysis (Dialyzers)
-
Principle: Diffusion across semipermeable membrane; blood and dialysate flow counter-current.
-
Components:
-
Dialyzer ("artificial kidney"): Hollow fibers (polysulfone); high surface area.
-
Blood Pump: Peristaltic.
-
Dialysate System: Concentrate mixed with water; ultrafiltration control.
-
Heparin Pump: Anticoagulation.
-
-
Clearance: Small solutes (urea, creatinine) diffuse; fluid removal via ultrafiltration (pressure gradient).
10. BIOTELEMETRY AND PATIENT MONITORING
10.1 Bio-Telemetry Systems
-
Block Diagram: Sensor → Signal Conditioning → Transmitter (modulator + RF) → Antenna → Receiver → Demodulator → Display/Recorder.
-
Types:
-
Wireless: RF (MHz–GHz), infrared; implantable (telemetry pill) vs. wearable.
-
Wired: For bed-bound patients (less common).
-
-
Advantages: Mobility, continuous monitoring, reduced infection risk (wired).
-
Disadvantages: Noise, interference, security (encryption needed), battery life.
10.2 Biometric Systems
-
Identification Methods:
-
Fingerprint: minutiae matching.
-
Iris: Pattern recognition (high accuracy).
-
Face: Feature extraction (geometry, texture).
-
Voice: Spectral analysis.
-
-
Applications: Patient ID, access control to records/areas, attendance.
10.3 Remote Patient Monitoring
-
Telemedicine Architecture: Wearable sensors → gateway (smartphone) → cloud/EMR → clinician dashboard.
-
Use Cases: Chronic disease (diabetes, hypertension), post-discharge, ICU (early warning scores).
11. EMBEDDED SYSTEMS IN BIOMEDICAL DEVICES
11.1 Embedded System Fundamentals
-
Definition: Dedicated computer system within larger device; real-time constraints.
-
Characteristics: Single-purpose, low power, high reliability, cost-sensitive.
-
Quality Attributes: Safety, maintainability, usability, real-time performance.
-
Design Metrics: Cost, performance (MIPS), power (mW), size, time-to-market.
-
Classification:
-
By Performance: Microcontroller (low), DSP (medium), ARM (high).
-
By Complexity: Small (8-bit), medium (16-bit), complex (32-bit with OS).
-
11.2 Processors for Embedded Biomedical Systems
| Type | Description | Example | Use |
|---|---|---|---|
| General-Purpose | Flexible, general instruction set | x86, ARM | High-end monitors |
| Application-Specific | Optimized for domain (e.g., DSP) | TMS320Cxx | ECG filtering, FFT |
| Single-Purpose | Hardwired logic; no program memory | ASIC, FPGA | Pulse oximeter algorithm |
-
RISC vs CISC:
| Feature | RISC | CISC | |---------|------|------| | Instruction Set | Fixed-length, few | Variable, many | | Execution | Single-cycle (pipelined) | Multiple cycles | | Registers | Many (16–32) | Few (8–16) | | Examples | ARM, MIPS, PIC | x86, 8051 |
-
Von Neumann vs Harvard:
-
Von Neumann: Single memory for code/data; simpler, slower (bus bottleneck).
-
Harvard: Separate memories/buses for code/data; faster, used in DSPs (e.g., TMS320).
-
11.3 Microcontroller 8051
-
Architecture:
-
ALU: 8-bit; operations on ACC, B.
-
Registers: ACC (A), B (for multiply/divide), PSW (flags), DPTR (16-bit data pointer), SP.
-
Memory: 128 B internal RAM (00–7F: general, 80–FF: SFRs); 4 KB internal ROM (code).
-
I/O Ports: P0–P3; quasi-bidirectional (pull-up resistors when written 1).
-
-
Timers/Counters:
-
Mode 0: 13-bit timer (THx: 8 bits, TLx: 5 bits).
-
Mode 1: 16-bit timer (THx, TLx full).
-
Mode 2: 8-bit auto-reload (TLx reloads THx on overflow).
-
Mode 3: T0 split into two 8-bit timers; T1 stopped.
-
Baud Rate (Serial Mode 1/3): $$\displaystyle BR = \frac{2^{SMOD}}{32} \times \frac{f_{osc}}{12 \times (256 - TH1)} $$.
-
-
Serial Communication:
-
Mode 0: Synchronous, shift register; baud = fₒₛc/12.
-
Mode 1: 8-bit UART; baud from Timer 1.
-
Mode 2: 9-bit UART; baud = fₒₛc/64 or /32 (SMOD).
-
Mode 3: 9-bit UART; baud variable (like Mode 1).
-
-
Interrupts: Sources: IE0 (external 0), TF0 (Timer 0), IE1, TF1, RI/TI (serial). Enable via IE register; priority via IP.
-
Interfacing:
-
ADC0808: 8-bit, 8-channel; start conversion on high-to-low pulse; EOC goes high when done; connect P1 for data, P2 for address.
-
DAC0832: 8-bit; two inputs (DAC0, DAC1); current output → op-amp for voltage.
-
Stepper Motor: Unipolar (5/6 wire) or bipolar (4 wire); sequence via port.
-
LCD 16x2: HD44780 controller; 4-bit or 8-bit mode; RS, RW, E signals.
-
-
Embedded C Example (Serial Mode 1):
#include <reg51.h> void main() { TMOD = 0x20; // Timer1, Mode2 (auto-reload) TH1 = 0xFD; // 9600 baud @ 11.0592 MHz SCON = 0x50; // Mode1, enable receive TR1 = 1; // Start Timer1 TI = 1; // Ready to transmit while(1) { SBUF = 'H'; while(!TI); TI=0; SBUF = 'E'; while(!TI); TI=0; SBUF = 'L'; while(!TI); TI=0; SBUF = 'L'; while(!TI); TI=0; SBUF = 'O'; while(!TI); TI=0; SBUF = '\n'; while(!TI); TI=0; } }
11.4 Microcontroller 8096
-
Functional Block Diagram:
-
CPU: 16-bit; 8-bit data bus, 16-bit address bus.
-
Memory: 4 KB ROM, 232 B RAM (internal); external up to 64 KB.
-
I/O: 4 8-bit ports (P0–P3); P0/P2 multiplexed with address/data.
-
Timers: Timer1 (16-bit, event counter), Timer2 (16-bit, watchdog).
-
ADC: 10-bit, 8-channel; successive approximation.
-
PWM: Pulse width modulation output.
-
Watchdog: Timer2; resets on overflow if not cleared.
-
Interrupt Controller: 5 sources (external, timer1, timer2, serial, ADC); priority levels.
-
-
Addressing Modes (with examples):
| Mode | Syntax | Example | Description | |------|--------|---------|-------------| | Immediate | LD R1, #05H |
LD R1, #05H| Load immediate 05H into R1 | | Direct | LD R1, 0200H |LD R1, 0200H| Load from memory 0200H | | Indirect | LD R1, @R2 |LD R1, @R2| Load from address in R2 | | Relative | JBC R1, label |JBC R1, LOOP| Jump if R1 bit clear | | Indexed | LD R1, 0200H[R2] |LD R1, 0200H[R2]| Load from 0200H + R2 | -
Instruction Set Classification:
-
Data Transfer:
LD,ST,PUSH,POP. -
Arithmetic:
ADD,SUB,MPY(multiply → R1:R0),DIV(divide R1 by R0). -
Logic:
AND,OR,XOR,NOT. -
Branch:
JMP,JBC,DJNZ(decrement and jump if not zero).
-
-
I/O Ports:
-
Structure: Port latch → buffer → pin. Configured as input (write 1 to latch) or output (write 0/1).
-
Modes: Input, output, bidirectional (P0/P2 for external memory).
-
-
Control/Status Registers:
-
CS0, CS1: Chip select for external memory.
-
PWM Control: Duty cycle register.
-
ADC Control: Start conversion, channel select.
-
-
Memory Mapping:
-
Internal: 0000–0FFFH (ROM), 1000–10FFH (RAM), 2000–20FFH (SFRs).
-
External: 8000–FFFFH (if CS0/CS1 configured).
-
11.5 ARM and DSP Processors
-
ARM:
-
RISC Features: Load-store architecture (memory access only via LDR/STR), fixed 32-bit instructions, large register file (R0–R15, CPSR).
-
Thumb: 16-bit compressed instruction set for code density.
-
Use: Mobile/wearable medical devices (low power, high performance).
-
-
DSP:
-
Harvard Architecture: Separate program/data memories/buses.
-
Hardware Multiplier: Single-cycle MAC (multiply-accumulate).
-
Pipelining: Deep pipelines for high throughput.
-
Circular Buffering: Auto-increment/decrement for FIR/IIR filters.
-
Applications: Real-time ECG/EEG filtering, FFT for spectral analysis.
-
11.6 Peripherals and Interfacing
-
Watchdog Timer:
-
Purpose: Recover from software hangs (e.g., infinite loop).
-
Implementation: Timer must be cleared periodically; if overflow, resets system.
-
Timing Diagram: Clear signal before timeout; timeout → reset pulse.
-
-
Interrupt Controller (8259A):
-
Priority Management: Fixed or rotating priority; cascade for >8 interrupts.
-
ISR Addressing: Vector provided by controller or CPU fetches from interrupt vector table.
-
-
Keyboard Controller (8279):
-
Modes:
-
Scanned: Keyboard scanned by CPU; debounce in software.
-
Interrupt: Key press generates interrupt; FIFO stores code.
-
-
Display Scanning: Multiplexed 7-segment/LED; refresh rate >50 Hz.
-
-
Display Interfacing:
-
7-segment: Common anode/cathode; multiplexed for multiple digits.
-
LCD (HD44780): Commands (clear, cursor home, entry mode); data write; 4-bit mode saves I/O.
-
-
RS-232:
-
Signals: TXD (transmit), RXD (receive), RTS (request to send), CTS (clear to send), DTR (data terminal ready), DSR (data set ready).
-
Handshaking: Hardware (RTS/CTS) or software (XON/XOFF).
-
Voltage Levels: ±3 to ±15 V; logic 1 = negative, 0 = positive.
-
11.7 Data Acquisition System Design
-
Block Diagram:
Sensor → Signal Conditioning (amp, filter) → Multiplexer → ADC → Microcontroller → (Storage/Display/Transmission) -
Hardware:
-
Op-amp: Low offset, low noise (e.g., INA128 for IA).
-
Filter: Anti-aliasing LPF (e.g., 150 Hz for ECG).
-
ADC: Resolution (12–16 bit), sampling rate (>2× max signal frequency).
-
Isolation: Opto-isolator or transformer for patient safety.
-
-
Software:
-
Sampling Rate: Nyquist: $$\displaystyle f_s > 2 f_{max} $$.
-
Processing: Digital filtering (FIR/IIR), calibration (offset/gain correction).
-
Example 8051 ECG Monitor:
-
Hardware: AD620 (IA), RC filter (0.05–150 Hz), ADC0808, LCD.
-
Software: Sample at 200 Hz, moving average filter, display heart rate.
-
-
12. POWER ELECTRONICS FOR MEDICAL EQUIPMENT
12.1 Power Semiconductor Devices
-
Power Diode:
-
Types: Standard (slow), fast recovery (reverse recovery <5 µs), Schottky (low Vf, low reverse voltage).
-
V-I: Forward: $$\displaystyle V_f \approx 0.7–1.2 $$ V; Reverse: blocking until breakdown.
-
-
Thyristor (SCR):
-
Structure: PNPN four-layer; three terminals (anode, cathode, gate).
-
Static Char: Forward blocking (J2 reverse-biased), forward conducting (latching current $$\displaystyle I_L $$ required to stay on), reverse blocking.
-
Dynamic: Turn-on time (delay + rise), turn-off time (reverse recovery + gate recovery).
-
Latching Current ($$\displaystyle I_L $$): Minimum anode current to maintain conduction after gate pulse.
-
Holding Current ($$\displaystyle I_H $$): Minimum anode current to keep SCR on.
-
-
TRIAC: Bidirectional SCR; 5-layer; triggering in any quadrant (I+, I-, III+).
-
DIAC: Two-terminal, bidirectional; breaks over at $$\displaystyle V_{BO} $$; used to trigger TRIACs.
-
Power MOSFET:
-
Structure: N-channel vertical; gate oxide insulation.
-
Transfer Char: $$\displaystyle I_D = k(V_{GS} - V_{th})^2 $$ (quadratic in saturation).
-
Switching Losses: $$\displaystyle E_{on/off} = \frac{1}{2} V_{DS} I_D t_{r/f} $$.
-
Applications: Low-voltage (<200 V), high-frequency (>100 kHz); DC-DC converters, battery chargers.
-
-
IGBT:
-
Structure: MOSFET gate + BJT output; P⁺ substrate, N⁻ drift, P body, N⁺ source.
-
V-I Char: MOSFET-like turn-on, BJT-like saturation (low $$\displaystyle V_{CE(sat)} $$).
-
Applications: Medium-power (600 V–3 kV); motor drives (MRI gradients), inverters.
-
12.2 AC-DC Converters (Rectifiers)
-
Single-Phase Half-Wave:
-
Output: $$\displaystyle V_{dc} = \frac{V_m}{\pi} $$ (uncontrolled); $$\displaystyle V_{dc} = \frac{V_m}{\pi}(1 + \cos\alpha) $$ (controlled).
-
Ripple frequency: $f$ (50 Hz).
-
-
Single-Phase Full-Wave:
-
Center-Tap: $$\displaystyle V_{dc} = \frac{2V_m}{\pi} $$ (uncontrolled); $$\displaystyle V_{dc} = \frac{2V_m}{\pi}\cos\alpha $$ (controlled).
-
Bridge (more common): $$\displaystyle V_{dc} = \frac{2V_m}{\pi} $$ (diodes); $$\displaystyle V_{dc} = \frac{2V_m}{\pi}\cos\alpha $$ (SCRs).
-
-
Three-Phase Full-Wave Bridge:
-
Diode: $$\displaystyle V_{dc} = \frac{3\sqrt{6}}{\pi} V_{LL} \approx 2.34 V_{LL} $$.
-
SCR: $$\displaystyle V_{dc} = \frac{3\sqrt{6}}{\pi} V_{LL} \cos\alpha $$.
-
-
RMS Output for Half-Controlled (Example):
Q: Single-phase full-wave half-controlled rectifier, $$\displaystyle V_m=300 $$ V, $$\displaystyle \alpha=30° $$. Find $$\displaystyle V_{rms} $$.
A: Output consists of positive half-sines from 30° to 180° and zero from 180° to 360°.
$$V_{rms} = \sqrt{\frac{1}{2\pi} \int_{30°}^{180°} (V_m \sin\theta)^2 d\theta} = V_m \sqrt{\frac{1}{2\pi} \left( \pi - \frac{\pi}{6} - \frac{\sin 60°}{2} \right)}$$
$$= 300 \sqrt{\frac{1}{2\pi} \left( \frac{5\pi}{6} - \frac{\sqrt{3}}{4} \right)} \approx 300 \times 0.577 = 173.1\ \text{V}$$
\boxed{V_{rms} = 173.1\ \text{V}}
12.3 DC-AC Converters (Inverters)
-
Single-Phase Bridge (180° Conduction):
-
Switches: T1/T4 on → +Vdc; T2/T3 on → -Vdc.
-
Resistive Load: Square wave output; $$\displaystyle V_{rms} = V_{dc} $$.
-
Inductive Load: Current continuous; output current sinusoidal (if filtered).
-
-
Three-Phase Bridge (120° Conduction):
-
Each switch conducts 120°; output line voltage: 6-step waveform.
-
$$\displaystyle V_{L,rms} = \sqrt{\frac{2}{3}} V_{dc} $$ (ideal).
-
-
PWM Inverter:
-
Principle: High-frequency carrier (e.g., 5 kHz) compared with sine reference; switches turned on/off to approximate sine.
-
Modulation Index $$\displaystyle m = V_{ref}/V_{car} $$; amplitude control.
-
Advantages: Voltage control without extra stage, harmonic reduction (carrier frequency harmonics).
-
-
Harmonic Reduction:
-
PWM: Shift carrier phase (phase-shifted PWM) to cancel low-order harmonics.
-
Multi-level: More voltage levels (e.g., NPC); lower dv/dt, THD.
-
Selective Harmonic Elimination: Solve nonlinear equations to eliminate specific harmonics.
-
12.4 AC Voltage Controllers
-
On-Off Control (Integral Cycle):
-
Whole cycles on/off; output $$\displaystyle V_{rms} = V_s \sqrt{k} $$ where $k$ = duty cycle (on cycles/total).
-
Low harmonic distortion, but discontinuous power.
-
-
Phase Control:
-
Firing angle $\alpha$ controls conduction angle; $$\displaystyle V_{rms} = V_s \sqrt{\frac{1}{2\pi} \int_{\alpha}^{\pi} \sin^2\theta d\theta} $$.
-
For RL load, must consider $$\displaystyle \phi = \tan^{-1}(\omega L/R) $$; conduction starts at $\alpha + \phi$ if $$\displaystyle \alpha < \pi - \phi $$.
-
Expression (RL load, continuous conduction):
-
$$V_{rms} = V_s \sqrt{\frac{1}{\pi} \left[ (\pi - \alpha) + \frac{\sin 2\alpha}{2} \right]}$$
- Applications: Light dimmers, heater control.
12.5 DC-DC Converters (Choppers)
-
Step-Down (Buck):
-
Switch (MOSFET) on: inductor charges (di/dt = (V_in - V_out)/L); diode reverse-biased.
-
Switch off: inductor discharges through diode to load (di/dt = -V_out/L).
-
V_out = D V_in (D = duty cycle = T_on/T); continuous if $$\displaystyle L > \frac{(1-D)R}{2f} $$.
-
-
Step-Up (Boost):
-
Switch on: inductor charges (di/dt = V_in/L); diode blocks.
-
Switch off: inductor voltage adds to V_in → V_out = V_in/(1-D).
-
-
Buck-Boost:
-
Inverting: $$\displaystyle V_{out} = -\frac{D}{1-D} V_{in} $$.
-
Non-inverting ( SEPIC ): $$\displaystyle V_{out} = \frac{D}{1-D} V_{in} $$.
-
-
Applications: Battery-powered devices (infusion pumps, portable monitors); input from battery (3.7 V Li-ion), output 5 V/3.3 V.
12.6 Cycloconverters
-
Principle: Direct AC-AC conversion; no DC link; output frequency $$\displaystyle f_o < f_i $$.
-
Single-Phase to Single-Phase:
-
Midpoint: Two antiparallel thyristor pairs; each half-cycle one pair conducts.
-
Bridge: Four thyristors; more flexible.
-
Operation: Firing angles $$\displaystyle \alpha_1 $$ (positive) and $$\displaystyle \alpha_2 $$ (negative) control output voltage.
-
-
Three-Phase: Used for large slow-speed motors (e.g., rolling mills, surgical tables).
-
Advantages: Regenerative (can feed back power), no intermediate DC link.
-
Limitations: Low output frequency (<½ input), complex control, poor power factor.
12.7 Switched-Mode Power Supplies (SMPS)
-
Principle: Switch at high frequency (20–500 kHz); transformer/inductor small; feedback loop regulates output.
-
Topologies:
-
Flyback (isolated):
-
Switch on: energy stored in transformer (primary current ramps up); secondary diode reverse-biased.
-
Switch off: energy transferred to secondary (diode conducts); output capacitor supplies load.
-
V_out = \frac{N_s}{N_p} \frac{D}{1-D} V_{in}.
-
-
Forward: Energy transferred while switch on; requires reset winding or tertiary winding.
-
Push-Pull: Two switches alternate; transformer flux bidirectional.
-
Half-Bridge: Two switches + capacitors; voltage across primary = V_in/2.
-
Full-Bridge: Four switches; full V_in applied.
-
-
Advantages over Linear: Efficiency >80% (vs. ~50%), smaller size, wide input range (85–265 VAC).
-
Applications: Universal input power supplies for all medical electronics (monitors, imaging systems).
12.8 Special Topics in Power Electronics
-
Commutation:
-
Natural (Line): AC source voltage reverses; SCR turns off when current crosses zero.
-
Forced:
-
External Pulse: Auxiliary SCR discharges capacitor into main SCR cathode.
-
Resonant: LC circuit creates zero current/voltage.
-
Complementary: Two SCRs in parallel with opposite polarity; one turns on to commutate other.
-
-
-
Series/Parallel Operation of Thyristors:
-
Series: Voltage sharing: use snubber (RC across each); derating factor (e.g., 2 in series → each rated for >½ total voltage).
-
Parallel: Current sharing: use small ballast resistors; match V-I characteristics.
-
String Efficiency:
-
$$\eta = \frac{\text{Total voltage/current rating}}{\text{Sum of individual ratings}} \times 100\%$$
> **Example**: 7.5 kV, 1 kA system; each SCR: 500 V, 75 A; derating 14% → effective rating: 430 V, 64.5 A.
> Series: $$\displaystyle N_s = \lceil 7500/430 \rceil = 18 $$; Parallel: $$\displaystyle N_p = \lceil 1000/64.5 \rceil = 16 $$; Total SCRs = 288; String efficiency = $$\displaystyle \frac{7500 \times 1000}{18 \times 500 \times 16 \times 75} \times 100\% \approx 55.6\% $$.
-
Effect of Source Inductance on Rectifiers:
-
Overlap Angle $\mu$: During commutation, both incoming and outgoing SCRs conduct.
-
Output Voltage Reduction: $$\displaystyle V_d = V_0 \cos(\alpha + \mu) $$ for three-phase full converter.
-
$$\displaystyle \mu \propto \frac{\omega L_s}{V_d} $$ (source inductance $$\displaystyle L_s $$).
-
-
Harmonic Control in Inverters:
-
Space Vector PWM: Better DC bus utilization, lower THD.
-
Selective Harmonic Elimination: Solve Fourier equations to set switching angles to eliminate specific harmonics (e.g., 5th, 7th).
-
13. SYSTEM INTEGRATION, SAFETY, AND STANDARDS
13.1 Medical Device Standards
-
IEC 60601-1: General requirements for basic safety and essential performance.
-
Protection Against Electric Shock:
-
Type B: Body (non-patient) contact; ≤100 µA leakage.
-
Type BF: Patient contact (except heart); ≤100 µA.
-
Type CF: Direct cardiac contact; ≤10 µA.
-
-
EMC: Emissions (CISPR 11), Immunity (IEC 61000-4 series).
-
-
IEC 62304: Medical device software life cycle processes.
13.2 Design for Reliability and Safety
-
Fault Tolerance:
-
Watchdog Timers: Reset on software hang.
-
Error Detection: Parity, CRC, checksums.
-
Redundancy: Triple modular redundancy (voting) for critical systems.
-
-
FMEA/FTA: Systematic analysis of failure modes and their effects; fault tree for top-event probability.
-
Software Safety: Coding standards (MISRA C), validation/verification, change control.
13.3 Testing and Validation
-
Electrical Safety: Ground resistance (<0.2 Ω), leakage current (per IEC 60601), dielectric strength (hi-pot test).
-
Performance: Accuracy (calibration against reference), repeatability, linearity.
-
Clinical Evaluation: Usability studies (human factors), clinical trials (efficacy/safety).
END OF UNIT 2 NOTES
Focus on past paper patterns: ECG/EEG, electrodes, 8051/8096 programming & architecture, power device characteristics, rectifier/inverter calculations, SMPS topologies, safety standards.