Skip to content
EX-303 · Electrical Measurements and Instruments/Quick Revision Short Notes

Electrical Measurements and Instruments (EX-303) - Unit 1 Short Notes

UNIT 1: ELECTRICAL MEASUREMENTS AND INSTRUMENTS


1.0 FUNDAMENTALS OF MEASUREMENT & ERROR ANALYSIS

1.1 Types of Errors in Measurement

  • Gross Errors: Human mistakes (reading, recording). Preventable by care and averaging.

  • Systematic Errors: Consistent, predictable. Sub-classified as:

    • Instrumental: Due to instrument limitations (e.g., calibration error, wear).

    • Environmental: Due to external conditions (temperature, humidity, magnetic field).

    • Observational: Due to observer's habit or parallax.

    • Can be minimized by calibration, correction factors, or improved technique.

  • Random Errors: Unpredictable variations (e.g., noise). Reduced by repeated measurements and statistical analysis.

[!TIP] EXAM TIP: Be prepared to classify a given error and suggest its minimization. Gross errors are not part of instrument specifications.

1.2 Static Characteristics of Instruments

  • Accuracy: Closeness to true value. (Absolute Error = Measured - True).

  • Precision: Repeatability of readings (clustering). High precision ≠ high accuracy.

  • Resolution: Smallest detectable change in input.

  • Sensitivity (Scale Factor): Output change per unit input change. For deflection instruments: S = dθ/dx (rad/unit). Static Sensitivity is slope of calibration curve.

  • Threshold: Minimum input to cause detectable output.

  • Hysteresis: Difference in output for increasing vs. decreasing input.

  • Linearity: Maximum deviation from ideal straight-line calibration curve.

  • Repeatability: Same conditions, same input → same output.

  • Reproducibility: Same input, different conditions (time, operator) → same output.

1.3 Dynamic Characteristics of Instruments

  • Concerned with response to time-varying inputs.

  • Speed of Response: How fast output follows input.

  • Lag: Delay in response (retardation type, time delay type).

  • Fidelity: How accurately output reproduces input waveform (shape).

  • Dynamic Error: Difference between dynamic output and true instantaneous value.

1.4 Loading Effects & Instrument Loading

  • Problem: Connecting a measuring instrument alters the circuit condition.

  • Voltmeter Loading: Voltmeter is connected in parallel. Its internal resistance (Rₘ) forms a shunt with the circuit resistance (Rₕ). The measured voltage is lower than the open-circuit voltage.

    Loading Error % = (V_loaded - V_true) / V_true × 100

    For high source resistance (Rₕ), use a voltmeter with high sensitivity (Ω/V) → high Rₘ.

  • Ammeter Loading: Ammeter is connected in series. Its internal resistance (Rₘ) adds to the circuit resistance. The measured current is lower than the true current.

    Use an ammeter with very low Rₘ.

  • Example: Multimeter on 10V scale with sensitivity 20,000 Ω/V has Rₘ = 200 kΩ. Loading a high-R circuit (e.g., Rₕ = 10 kΩ) causes significant error.

[!TIP] COMMON PITFALL: Always check instrument impedance relative to circuit impedance. For voltage, Rₘ >> Rₕ. For current, Rₘ << Rₕ.


2.0 CLASSIFICATION & OPERATING PRINCIPLES OF ANALOG INSTRUMENTS

2.1 Classification of Analog Instruments

  • By Operating Principle:

    • Electromagnetic: PMMC, Moving Iron (MI), Electrodynamic (Dynamometer).

    • Electrostatic: For high voltage AC/DC.

    • Thermal: For RMS of any waveform (thermocouple, thermistor).

    • Electrochemical: pH meters, conductivity.

  • By Quantity Measured: Ammeters, Voltmeters, Wattmeters, Energy Meters, etc.

2.2 Permanent Magnet Moving Coil (PMMC) Instruments

  • Construction: Permanent magnet (high coercivity steel), cylindrical air gap, moving coil (many turns, aluminum former), spring control (hair-springs), eddy current damping (aluminum frame in magnetic field), pointer, scale.

  • Principle: Current in coil → magnetic field interaction with permanent magnet → torque T_d = BINA (B=flux density, I=current, N=turns, A=area). Linear scale because T_d ∝ I.

  • Torque Equation: T_d = k₁ I (k₁ = BNA).

  • Control Torque: T_c = k₂ θ (spring constant).

  • Equilibrium: T_d = T_c → θ ∝ I.

  • Applications:

    • Ammeter: Low internal resistance. Shunt required for range extension.

    • Voltmeter: High internal resistance. Series multiplier required.

  • Limitations: AC measurement requires rectifier. Sensitive to overload, position.

2.3 Moving Iron (MI) Instruments

  • Construction Types:

    • Attraction Type: Single iron piece, coil. Current → iron attracted to coil → movement.

    • Repulsion Type: Two iron vanes (one fixed, one movable) inside coil. Current → both magnetized similarly → repulsion → movement.

  • Principle: Non-linear torque. T_d ∝ I² (for both AC & DC). Hence, scale is non-linear (crowded at lower end).

  • Torque Equation (Repulsion Type): T_d ∝ (N₁I)² (N₁ = turns on fixed coil). Works for AC/DC without polarity.

  • Damping: Air friction damping (vane in air chamber).

  • Advantages: Robust, cheap, can measure AC/DC, high overload capacity.

  • Limitations: Non-linear scale, low sensitivity, susceptible to stray fields (needs shielding).

2.4 Electrodynamic (Dynamometer) Instruments

  • Construction: Two coils:

    • Fixed Coils (Current Coil): Carry current to be measured. Split to allow moving coil movement.

    • Moving Coil (Pressure/Voltage Coil): Carries current proportional to voltage (via series multiplier). Air-cored (no iron).

  • Principle: Interaction of magnetic fields from both coils. Torque exists for both AC & DC.

  • Torque Equation (AC/DC): T_d = k I₁ I₂ cos φ (I₁ = current coil current, I₂ = pressure coil current, φ = phase angle). For DC, φ=0, T_d ∝ I₁ I₂.

  • Scale: Non-linear (√ type) because T_d ∝ I₁ I₂ and for voltmeter/ammeter configurations, one current is fixed or proportional to the other.

  • Applications:

    • Wattmeter: (I₁ = load current, I₂ ∝ load voltage).

    • Ammeter/Voltmeter (less common due to low sensitivity).

  • Damping: Eddy current damping (moving coil on aluminum frame).

  • Special Features for Low Power Factor Wattmeter:

    • High torque design: To overcome low cos φ torque.

    • Compensated pressure coil circuit: Series capacitor to make pressure coil circuit resistive (minimize phase error).

    • Air damping: To avoid hysteresis errors from fluid damping at low PF.

    • Light moving system: For high sensitivity.


3.0 INSTRUMENT TRANSFORMERS (CT & PT)

3.1 Current Transformer (CT)

  • Construction: Primary is single turn (conductor or bar). Secondary is multi-turn wound on laminated core. Secondary rated current usually 5A or 1A. Burden = external impedance (Z_b).

  • Equivalent Circuit & Phasor Diagram: Similar to a step-down transformer under short-circuited secondary. Key: Exciting current (I₀) has two components:

    • Magnetizing component (Iₘ): Creates core flux.

    • Core loss component (Iₑ): Compensates core loss.

  • Ratio Error: n = (Kₙ Iₛ - I₁) / I₁ × 100% (Kₙ = nominal ratio I₁ₙ/Iₛₙ). Caused by exciting current.

  • Phase Angle Error (θ): Angle between primary current (I₁) and secondary current reversed (I₂').

  • Errors & Minimization:

    • Effect of Exciting Current: Increases ratio error. Minimized by using high permeability core, low flux density.

    • Effect of Secondary Burden (Z_b): Higher Z_b → larger voltage drop in secondary → larger I₀ → larger errors. Keep burden within rated VA.

    • Effect of PF of Primary Current: Affects phase angle error significantly.

  • ⚠️ DANGER: Open-Circuited Secondary: No secondary current → all primary current becomes exciting current → extremely high core flux → dangerously high voltage across open secondary terminals → insulation breakdown, safety hazard. Never open CT secondary when primary is energized.

  • Testing: Ratio test (compare primary/secondary currents), phase angle test (using zero-phase-factor detector).

3.2 Potential Transformer (PT)

  • Construction: Shell-type or core-type. Primary has many turns (for high voltage), secondary few turns (usually 100-110 V). Designed for high accuracy at near-rated burden.

  • Equivalent Circuit & Phasor Diagram: Similar to a step-down transformer. Exciting current (I₀) drawn by magnetizing branch.

  • Ratio Error: Caused by voltage drop in windings (I₁R₁, I₁X₁, I₂R₂, I₂X₂) and I₀.

  • Phase Angle Error: Angle between primary voltage (V₁) and reversed secondary voltage (V₂').

  • Errors & Minimization:

    • Effect of Voltage Drop: Minimize by using thick conductors (low R, X).

    • Effect of Burden & its PF: Higher burden or low PF burden → larger voltage drops → larger errors. Operate near rated burden.

    • Effect of Magnetizing Current: Use high permeability core, low flux density.

  • Testing: Ratio test (apply known V₁, measure V₂), phase angle test.

3.3 Comparison of CT and PT

Feature Current Transformer (CT) Potential Transformer (PT)
Primary Connection In series with load. In parallel with load.
Primary Current Determined by load. Determined by system voltage & impedance.
Secondary Current Nearly constant (rated). Nearly constant (rated) at rated burden.
Operating Condition Nearly short-circuited (low Z_b). Nearly open-circuited (high Z_b).
Dangerous Condition Open secondary (high voltage). Short secondary (high current).
Core Flux Density Low (designed for low VA). Higher (designed for accuracy).
Primary Turns 1 (or few). Many.

4.0 POWER MEASUREMENT - WATTMETERS

4.1 Single-Phase Electrodynamometer Wattmeter

  • Construction: Fixed coils (current coil, series with load). Moving coil (pressure coil, series with high resistance multiplier, parallel to load).

  • Working Principle: T_d ∝ (Flux from CC) × (Flux from PC) × cos(angle between them).

    • CC flux φ₁ ∝ I₁ (load current).

    • PC flux φ₂ ∝ I₂ (current ∝ load voltage V).

    • Angle between fluxes ≈ phase angle (φ) between V and I (since PC circuit made resistive).

    • Hence, T_d ∝ V I cos φ → Deflection ∝ Real Power.

  • Phasor Diagram: Shows V, I, and fluxes. PC current I₂ lags V by small angle α (due to PC inductance).

  • Scale: Non-linear (√ type).

  • Errors in Electrodynamometer Wattmeter:

    • Stray Magnetic Field Errors: External fields distort CC/PC fluxes → error. Minimized by shielding (iron case).

    • Errors due to Inductance/Capacitance of Pressure Coil: I₂ lags V → T_d ∝ V I cos(φ+α). Error = VI sin φ sin α. Compensated by adding a capacitor in parallel with PC to make PC circuit resistive.

    • Friction, Hysteresis, Eddy Currents: Standard mechanical/electrical errors.

4.2 Special Types of Wattmeters

  • Two-Element Wattmeter: Two separate dynamometer elements mounted on same spindle. Used for 3-phase, 3-wire (without neutral). Each element measures power in one phase.

  • Three-Element Wattmeter: Three separate elements. Used for 3-phase, 4-wire (with neutral). Each element measures phase power.

  • Low Power Factor Wattmeter:

    • High Torque: Light moving system, strong permanent magnets.

    • Compensated Pressure Coil Circuit: Series capacitor to neutralize PC inductance → make PC current in phase with voltage.

    • Air Damping: Avoids fluid hysteresis errors at low PF.

    • Reduced Control Spring Torque: Allows full-scale deflection at low cos φ.

4.3 Three-Phase Power Measurement

  • Two-Wattmeter Method:

    • Circuit: Two wattmeters connected in any two lines (say L1, L2). Current coils in two lines, pressure coils from respective lines to third line (L3).

    • Proof (Vector Diagram): W₁ = V₁I₁ cos(30°+φ), W₂ = V₂I₂ cos(30°-φ). For balanced load, V₁=V₂=V_L, I₁=I₂=I_L.

    • Total Power: P = W₁ + W₂ (Algebraic sum).

    • Condition for One Wattmeter Negative: φ > 60° (i.e., PF < 0.5 lagging or > 0.5 leading). One wattmeter reads negative.

  • Three Wattmeters: Used for 3-phase, 4-wire (unbalanced). P = W₁ + W₂ + W₃.


5.0 ENERGY MEASUREMENT - ENERGY METERS

5.1 Single-Phase Induction (Kilo-Watt-Hour) Meter

  • Construction:

    1. Driving System: Two electromagnets (Voltage Magnet - shunt, Current Magnet - series). Shading band on voltage magnet.

    2. Moving System: Aluminum disc on spindle, in air gap of magnets.

    3. Braking System: Permanent magnet (disc rotates in its field) → eddy currents → braking torque T_b ∝ ω (speed).

    4. Registering System: Gear train & dials (cyclometer or pointer type).

  • Working Principle:

    • Voltage flux φ_v (from V magnet, lagging V by ~90° due to shading).

    • Current flux φ_i (from I magnet, in phase with I).

    • Eddy currents induced in disc by φ_v interact with φ_i → driving torque T_d ∝ φ_v φ_i sin θ ∝ V I cos φ.

    • Braking torque T_b ∝ ω.

    • Steady speed: T_d = T_b → ω ∝ V I cos φ → Disc speed ∝ Power.

    • Energy: E = ∫ P dt ∝ ∫ ω dt ∝ Number of revolutions (N).

    • Meter Constant (K): K = N / E (rev/kWh).

  • Phasor Diagram: Shows φ_v (lagging V), φ_i (in phase with I), and resultant torque.

  • Errors in Energy Meter:

    • Friction: At light loads, T_d may not overcome friction → under-registration.

    • Creeping: Disc rotates with voltage only (no load) due to overcompensation, friction, or stray fluxes. Test: Creep test (V applied, I=0).

    • Braking Magnet: Weak magnet → high speed (under-registration). Strong magnet → low speed (over-registration).

    • Voltage, Frequency, Power Factor Errors: Due to non-ideal characteristics (shading, phase shifts).

    • Self-Braking & Stray Torques: Imperfect design.

  • Testing & Calibration:

    • Full Load Test:

      % Error = (Actual Rev - Expected Rev) / Expected Rev × 100

      Expected Rev = (P × t) / (3600 × K) (P in kW, t in sec, K in rev/kWh).

    • Light Load Test: Check friction/creeping.

    • Power Factor Test: At lagging/leading PF.

    • Example (Jun 2025): 50A, 230V meter, K=520 rev/kWh, makes 61 rev in 37 sec.

      P_true = V×I = 230×50 = 11500 W = 11.5 kW

      Expected Rev = (11.5 × 37) / (3600 × 520) = 425.5 / 1872000? Correct Formula: Expected Rev = (P (kW) × t (hr)) × K. t = 37/3600 hr. Expected Rev = 11.5 × (37/3600) × 520 = 61.8 rev. % Error = (61 - 61.8)/61.8 × 100 = -1.3%.

5.2 Three-Phase Energy Meter

  • Construction: Either single element with multiple discs (for 3-phase, 3-wire) or multiple elements (two-element for 3-wire, three-element for 4-wire).

  • Connection:

    • 3-phase, 3-wire: Two elements, each connected between a line and the other two lines (like two-wattmeter method). Total energy = sum of both elements.

    • 3-phase, 4-wire: Three elements, each between a line and neutral.

5.3 Maximum Demand Meter

  • Principle: Measures average power over a fixed interval (e.g., 15 or 30 min). Records the highest such average demand in the billing period.

  • Construction: Often a motor-driven integrating meter with a resetting mechanism (thermal or clock-driven). Pointer indicates current demand; maximum demand pointer is latched and resets manually/monthly.

5.4 Digital Electronic Energy Meter

  • Block Diagram:

    
    Voltage Sensor (Potential Divider) → ADC → Microcontroller/Processor ← Current Sensor (CT/Shunt)
    
                                                              ↓
    
                                                          Energy Calculation (∫VIdt)
    
                                                              ↓
    
                                                          Display (LCD/LED)
    
                                                              ↓
    
                                                          Memory/Output (Pulse, RS485)
    
    
  • Principle: Samples V & I, multiplies instantaneously, integrates over time. Highly accurate, multiple tariffs, tamper-proof.


6.0 SPECIAL PURPOSE INSTRUMENTS & MEASUREMENT TECHNIQUES

6.1 Ballistic Galvanometer

  • Construction: Moving coil instrument with very large moment of inertia (heavy coil) and very weak control (long period, T > 5 sec). Critical damping is avoided (underdamped).

  • Principle: Measures total charge (Q) passed as a current pulse. The coil receives an impulse → oscillates. First maximum deflection (θ₁) ∝ Q.

  • Equation of Motion (for first swing):

    Q = √(4πC K) × θ₁ (for critically damped case, but for BG, damping is small).

    More generally, Q = (C √(K/J)) × θ₁ where C = damping constant, K = control constant, J = inertia.

    Key: θ₁ ∝ Q if damping is constant.

  • Calibration: Often done using a known capacitor charged to known voltage.

  • Uses: Measure charge, capacitance, magnetic flux linkage.

  • Comparison with Flux Meter:

    | Feature | Ballistic Galvanometer | Flux Meter | | :--- | :--- | :--- | | Moving System | Heavy coil, high inertia | Light coil, low inertia | | Damping | Very light (underdamped) | Heavy damping (dead beat) | | Deflection | Proportional to total charge (Q) | Proportional to flux linkage (dΦ/dt) | | Response | Slow (periodic) | Fast, direct reading | | Measurement | Charge, capacitance, flux (via dΦ/dt) | Direct flux measurement |

6.2 Measurement of Magnetic Properties (B-H Curve)

  • Method of Reversals:

    1. Specimen (ring or bar) wound with primary (N₁) and secondary (N₂) windings.

    2. Apply slowly increasing DC to primary. At each current I, reverse primary current rapidly.

    3. Integrating voltmeter (ballistic galvanometer) connected to secondary measures ∫e dt = N₂ A ΔB (A = cross-section).

    4. Ballistic deflection θ ∝ ΔB. Plot θ vs. I (∝ H) → hysteresis loop.

    5. Advantage: Eliminates effect of residual magnetism and eddy currents.

  • Six-Point Method:

    1. Apply known currents (I₁, I₂, ... I₆) to primary (H ∝ I).

    2. For each I, measure secondary EMF e = N₂ A dB/dt with fluxmeter (not BG).

    3. ΔΦ = ∫ e dt between two points gives ΔB.

    4. Plot B vs. H point by point → dc hysteresis loop.

  • Lloyd-Fischer Square Method:

    • Purpose: Measure iron loss (hysteresis + eddy) in a specimen.

    • Setup: Square specimen (lamination) placed in two-coil system (primary for excitation, secondary for induced EMF). Secondary connected to wattmeter.

    • Principle: Wattmeter reading = Core loss × Volume. By knowing dimensions and frequency, calculate core loss per kg.

6.3 Measurement of Resistance

  • Wheatstone Bridge:

    • Circuit: Four arms (R₁, R₂, R₃, Rₓ). Galvanometer (G) between junctions, battery across other two.

    • Balance Condition: R₁/R₂ = R₃/Rₓ → Rₓ = R₃ × (R₂/R₁).

    • Used for: Medium resistances (1 Ω to 1 MΩ).

    • Sources of Error:

      • Contact resistance & lead resistance: In series with low Rₓ → error. Minimized by using four-terminal (Kelvin) connections for Rₓ.

      • Thermal EMFs: Dissimilar metal junctions in bridge → stray voltages. Use same metal, reverse battery, average.

      • Bridge sensitivity: Poor galvanometer sensitivity.

  • Kelvin's Double Bridge (Thomson Bridge):

    • Purpose: Low resistances (< 1 Ω) where lead/contact resistance matters.

    • Construction: Two pairs of ratio arms (P,Q and p,q). Unknown low resistance (Rₓ) and standard (Rₛ) are four-terminal (separate current and potential leads).

    • Principle: Potential leads of Rₓ and Rₛ connected to points that eliminate effect of lead/resistance (r) of Rₓ and Rₛ.

    • Balance Condition Derivation:

      At balance (G=0), potential at a = potential at c.

      I₁P = I₂p and I₁Q = I₂q → P/Q = p/q = m (say).

      Also, I₁(R + r₁) + I₂Rₛ = I₁Rₓ + I₂(r₂ + r).

      Using current ratios, solving gives:

      Rₓ = Rₛ × (P/Q) + (q/Q) × (r - m r₁)

      If P/Q = p/q (i.e., m = p/q), then Rₓ = Rₛ × (P/Q). The error term vanishes.

    • Superiority: Four-terminal Rₓ, Rₛ eliminate their own lead/contact resistances. Second set of ratio arms cancels remaining lead resistance (r).

  • Megger (Insulation Tester):

    • Principle: Hand-cranked or battery-operated high voltage DC source (500V, 1000V, 2500V). Measures very high resistances (MΩ to GΩ).

    • Use: Test insulation resistance of cables, motors, transformers. "Megger" is a trade name, now generic.

6.4 Measurement of Earth Resistance

  • Fall-of-Potential (Three-Point) Method:

    1. Earth electrode (E) under test.

    2. Two auxiliary electrodes (P, C) driven in earth, in a straight line, spaced ≥ 20m.

    3. Apply known current I between E and C.

    4. Measure voltage V between E and P.

    5. R = V/I. Move P to find position where V is maximum → that's true earth resistance.

  • Four-Point Method: Uses two current and two potential electrodes, eliminates effect of connecting lead resistances.

  • Earth Tester: Portable instrument with hand-driven generator, voltage/current meters, and terminals for electrodes. Shows direct reading.

6.5 Frequency Meters

  • Resonant (Vibrating Reed): Tuned reeds resonate at specific frequencies. Visual indication.

  • Weston (Electrodynamometer) Type: Two fixed coils (reactive & resistive), moving coil. Deflection ∝ frequency.

  • Digital Frequency Meter:

    • Block Diagram:

      
      Signal → Shaper (Schmitt trigger) → Gate (controlled by time base) → Counter → Display
      
      
    • Principle: Counts number of cycles of input signal in a precise time interval (1 sec, 0.1 sec) generated by crystal oscillator. f = Count / Gate time.

6.6 Ratio Meters

  • Purpose: Measure ratio of two quantities (e.g., voltage ratio in PT, current ratio in CT, impedance ratio).

  • Principle: Often use two dynamometer elements on same shaft. One element's torque ∝ I₁², other's ∝ I₂². Net deflection ∝ (I₁² - I₂²) or difference. Balanced when I₁ = I₂.

  • Use: CT/PT testing, impedance measurement.


7.0 INSTRUMENT DAMPING, RANGE EXTENSION & CALIBRATION

7.1 Damping Methods

  • Purpose: Bring pointer to rest quickly without oscillation.

  • Air Friction Damping: Piston in air chamber (PMMC, MI). Works in any position.

  • Eddy Current Damping: Most common. Moving conductor (aluminum frame) in magnetic field → eddy currents → opposing torque. Used in PMMC, Dynamometer. Requires permanent magnet.

  • Fluid Friction Damping: Vane in oil. High damping, but position-sensitive, messy.

  • Electromagnetic (Electrical) Damping: Short-circuiting moving coil (in PMMC) or using a separate damping coil.

7.2 Range Extension of Instruments

  • Ammeter Range Extension (Shunt):

    Let I = full-scale current, Iₘ = meter current, Rₘ = meter resistance.

    Shunt Rₛ carries Iₛ = I - Iₘ.

    Rₛ = (Iₘ Rₘ) / (I - Iₘ)

    Multi-range ammeter: Multiple shunts with rotary switch.

  • Voltmeter Range Extension (Series Multiplier):

    Let V = full-scale voltage.

    Rₛ must drop V - Vₘ where Vₘ = Iₘ Rₘ.

    Rₛ = (V - Vₘ) / Iₘ = (V/Iₘ) - Rₘ

    Multi-range voltmeter: Multiple series resistors with rotary switch. Use high-voltage multiplier for very high ranges.

7.3 Testing & Calibration of Instruments

  • Procedure: Compare instrument reading with a standard of higher accuracy (potentiometer, standard cell, calibrated reference meter).

  • Errors Determination:

    • Absolute Error (e): e = X_measured - X_true

    • Relative Error (ε): ε = e / X_true

    • Percentage Error: %e = (e / X_true) × 100 or (e / Full Scale) × 100 (for instruments).

  • Example (Calibration): Use a potentiometer to provide precise voltage/current. Compare readings. Plot calibration curve (error vs. reading).

  • Ammeter/Voltmeter Calibration: Use standard shunt/resistor and precise voltmeter to measure voltage drop across known resistance → calculate true current. Or use potentiometric method.

[!TIP] EXAM TIP: Always specify whether % error is based on true value or full-scale deflection (FSD). In energy meter full-load test, it's based on expected revolutions (true power).

Go to where you left off?

Quick Add to Notes

Save questions, your own notes and screenshots into notes filed by unit. It takes a free account.

Create free account

Have an account? Log in