1.0 FOUNDATIONS OF MEASUREMENT SCIENCE
1.1 Elements of a Generalized Measurement System
A measurement system typically consists of four functional stages:
-
Primary Sensing Element / Transducer: Converts the measured physical quantity (measurand) into a usable electrical/mechanical signal (e.g., thermocouple, strain gauge).
-
Signal Conditioning / Transmitter: Modifies the primary signal into a suitable form (amplification, filtering, linearization, conversion to standard signal like 4-20 mA).
-
Signal Processing / Data Handling: Further manipulates the signal (computation, correction, digitization, communication).
-
Display / Recording / Output: Presents the measurement result to the user (digital display, chart recorder, computer interface).
[!TIP] Exam Focus: Be prepared to map a specific instrument (e.g., pressure gauge, temperature controller) to these four elements.
1.2 Measurement Errors & Uncertainties
Definitions:
-
Error: The difference between the measured value and the true value.
-
Systematic Error: Consistent, repeatable deviation (calibration error, zero error, environmental effect). Correctable.
-
Random Error: Unpredictable fluctuations (noise, observer variation). Reducible by averaging.
-
Gross Error: Mistake (reading error, recording error). Eliminated by care.
-
-
Uncertainty: A quantitative estimate of the doubt about the measurement result. Expressed as:
-
Absolute Uncertainty (±Δx): Same units as measurement.
-
Relative/Percentage Uncertainty: $$\displaystyle \frac{\Delta x}{x} \times 100\% $$.
-
Sources of Errors:
-
Instrumental: Imperfect calibration, hysteresis, wear.
-
Environmental: Temperature, pressure, humidity, vibration changes.
-
Observational: Parallax, estimation error.
-
Methodological: Inherent limitations of the measurement method.
Key Calculation: Accuracy Specification
-
% of Full Scale Deflection (FSD): $$\displaystyle \text{Error} = \pm (\% \text{ of FSD}) $$. Absolute error is constant across range.
- Example (DEC 2024): Gauge range = 1000 kN/m², accuracy ±1% FSD → Max error = ±10 kN/m². At 100 kN/m², reading = $100 \pm 10$ kN/m².
-
% of True Value (Reading): $$\displaystyle \text{Error} = \pm (\% \text{ of reading}) $$. Absolute error varies with reading.
- Example (DEC 2024): Same gauge, ±1% of true value → At 100 kN/m², max error = ±1 kN/m². Reading = $100 \pm 1$ kN/m².
Mitigation: Regular calibration, averaging readings, environmental control, proper training.
[!TIP] Common Pitfall: Confusing %FSD (worse at low readings) with %True Value (constant relative error). Always check how instrument accuracy is specified.
1.3 Calibration & Standards
-
Need: To establish traceability to national/international standards, ensure accuracy, and determine system equation (output vs. input).
-
Standards Hierarchy:
-
Primary Standard: Highest accuracy, maintained by national labs (e.g., NPL India).
-
Secondary Standard: Calibrated against primary, used in industry labs.
-
Working Standard: Used routinely for calibrating instruments.
-
-
Procedure:
-
Apply known, precise inputs (from a standard) across the instrument's range.
-
Record corresponding outputs.
-
Plot calibration curve (output vs. input).
-
Determine system equation (often linear: $$\displaystyle y = mx + c $$) and correction factors.
-
-
Dead Weight Tester (Pressure - JUN 2025):
-
Principle: $$\displaystyle P = \frac{F}{A} = \frac{mg}{A} $$. Known masses (m) on a precision piston (area A) generate exact pressure.
-
Construction: Piston-cylinder assembly (low friction, high finish), fluid medium, weight set.
-
Role: Primary standard for pressure calibration.
-
1.4 Instrument Classification & Selection
-
By Function: Indicating (dial), Recording (chart), Integrating (totalizer).
-
By Output: Analog (continuous), Digital (discrete), Smart (digital with communication).
-
Key Specifications for Selection (from Datasheet):
-
Range & Span: Operating limits (Span = Max - Min).
-
Accuracy: % of reading or %FSD.
-
Resolution: Smallest detectable change.
-
Precision/Repeatability: Closeness of repeated readings.
-
Linearity: Max deviation from straight line.
-
Hysteresis: Difference between up-scale and down-scale readings for same input.
-
Deadband: Range of input change with no output change.
-
Drift: Change in output over time for constant input (zero/sensitivity drift).
-
2.0 STATIC PERFORMANCE CHARACTERISTICS (No time variation)
Definitions & Significance:
| Characteristic | Definition | Significance |
|---|---|---|
| Accuracy | Closeness to true value. | Overall quality of measurement. |
| Precision | Closeness of agreement among repeated measurements. | Repeatability/reproducibility. |
| Repeatability | Precision under same conditions (short-term). | Instrument stability. |
| Reproducibility | Precision under changed conditions (different operator, time). | Robustness. |
| Sensitivity | Ratio of output change to input change ($$\displaystyle S = \frac{\Delta \text{output}}{\Delta \text{input}} $$). | Responsiveness. |
| Scale Factor | Inverse of sensitivity. | |
| Threshold | Minimum input to produce detectable output change. | Detectability limit. |
| Resolution | Smallest input change that causes a detectable output step. | Discreteness of measurement. |
| Hysteresis | Difference in output for same input depending on direction (up/down). | Memory effect, friction. |
| Linearity | Max deviation from ideal straight line through calibration points. | Ease of calibration/use. |
| Span/Range | Operating limits (e.g., 0-100°C). | Application suitability. |
| Drift | Output change over time for constant input. | Long-term stability. |
| Fidelity | Ability to reproduce input waveform shape (often dynamic). | Accuracy for varying signals. |
[!TIP] Exam Link: These characteristics directly contribute to total measurement uncertainty. A question may ask: "How do hysteresis and nonlinearity impact accuracy?"
3.0 DYNAMIC PERFORMANCE CHARACTERISTICS & SIGNAL ANALYSIS
3.1 Dynamic Response Fundamentals
Needed when input varies with time. Instruments have inertia/lag, so output doesn't instantly track input. Characterized by time domain (response vs. time) and frequency domain (response vs. frequency).
3.2 System Modeling & Differential Equations
-
Lumped Parameter Models: Assume physical properties concentrated at a point.
-
First-Order System (e.g., thermometer, RC circuit):
$$ \tau \frac{dy}{dt} + y = Kx $$
Where:
* $y$ = output, $x$ = input
* $K$ = static sensitivity (steady-state gain)
* $\tau$ = **time constant** (time to reach 63.2% of final value for step input).
- Second-Order System (e.g., spring-mass-damper, RLC circuit):
$$ \frac{d^2y}{dt^2} + 2\zeta\omega_n\frac{dy}{dt} + \omega_n^2 y = K\omega_n^2 x $$
Where:
* $$\displaystyle \omega_n $$ = **natural frequency** (rad/s)
* $\zeta$ = **damping ratio** (dimensionless)
* Response type: Underdamped ($$\displaystyle \zeta<1 $$), Critically damped ($$\displaystyle \zeta=1 $$), Overdamped ($$\displaystyle \zeta>1 $$).
3.3 Time Domain Response
-
Standard Test Inputs: Step (sudden change), Ramp (linear increase), Impulse (very short pulse), Parabolic.
-
First-Order to Step Input ($$\displaystyle x(t)=X_0u(t) $$):
$$ y(t) = KX_0 (1 - e^{-t/\tau}) $$
* Reaches **98%** of final value at $t \approx 4\tau$, **99.3%** at $$\displaystyle t=5\tau $$.
-
Second-Order to Step Input ($$\displaystyle \zeta<1 $$):
-
Rise Time ($$\displaystyle t_r $$): Time to go from 10% to 90% of final value.
-
Peak Time ($$\displaystyle t_p $$): Time to first maximum overshoot.
-
Maximum Overshoot ($$\displaystyle M_p $$): $$\displaystyle M_p = e^{\frac{-\zeta\pi}{\sqrt{1-\zeta^2}}} \times 100\% $$.
-
Settling Time ($$\displaystyle t_s $$): Time to stay within a band (e.g., ±2% → $$\displaystyle t_s \approx \frac{4}{\zeta\omega_n} $$).
-
Damped Natural Frequency ($$\displaystyle \omega_d $$): $$\displaystyle \omega_d = \omega_n\sqrt{1-\zeta^2} $$.
-
3.4 Frequency Domain Analysis
- Fourier Transform (FT): Decomposes a time-domain signal $x(t)$ into its frequency components $X(f)$.
$$ X(f) = \int_{-\infty}^{\infty} x(t) e^{-j2\pi ft} dt $$
* **Significance (JUN 2025, NOV 2023):** Analyzes how a system responds to sinusoidal inputs of different frequencies. Essential for understanding **bandwidth** and **filtering**.
* **Property (NOV 2023):** **Linearity:** FT of a sum is sum of FTs: $$\displaystyle \mathcal{F}\{a x_1(t) + b x_2(t)\} = a X_1(f) + b X_2(f) $$.
-
Frequency Response: System's steady-state output to sinusoidal input $$\displaystyle x(t)=X_0\sin(\omega t) $$. Output: $$\displaystyle y(t)=Y_0\sin(\omega t + \phi) $$.
-
Magnitude Ratio: $$\displaystyle M(\omega) = \frac{Y_0}{X_0} $$.
-
Phase Angle: $\phi(\omega)$.
-
Plotted as Bode plots (log-log magnitude, log-linear phase).
-
-
Bandwidth: Frequency range where $M(\omega)$ is within -3 dB (or 70.7%) of low-frequency gain.
3.5 Input Signal Classification
-
Periodic Signals: Repeat after period $T$ (e.g., sine, square). Can be harmonic (single frequency sine) or non-harmonic periodic (complex waveform, requires Fourier series).
-
Aperiodic/Random Signals: Non-repeating, described statistically (mean, RMS, probability density). Response analysis uses statistical methods (JUN 2024).
3.6 Dynamic Performance Specs
-
Fidelity: Ability to reproduce input waveform shape accurately. High for systems with wide bandwidth.
-
Speed of Response: How quickly system reaches steady-state (e.g., time constant $\tau$, settling time $$\displaystyle t_s $$).
-
Measuring Lag / Time Lag (NOV 2023): Delay between input change and observable output change. Inversely related to speed.
-
Bandwidth: Higher bandwidth → faster response, better fidelity for rapid inputs.
4.0 SENSORS & TRANSDUCERS BY MEASURAND
4.1 Displacement, Position & Angular Measurements
-
Potentiometer (JUN 2025):
-
Principle: Variable resistance voltage divider. Displacement moves wiper on resistive element.
-
Construction: Wire-wound (resolution limited by wire turns), carbon film (continuous).
-
Conversion: Linear/rotary displacement → Voltage/current.
-
Limitations: Loading effect (needs high impedance buffer), mechanical wear, limited frequency response, finite resolution.
-
DiagramSEARCH: "potentiometer construction working principle"
-
-
LVDT & RVDT (JUN 2025 - 14m):
-
LVDT (Linear):
-
Construction: One primary winding, two identical secondary windings (series opposing), movable ferromagnetic core.
-
Principle: AC excitation on primary. Core position induces voltages in secondaries. Null position (core centered) → equal & opposite secondary voltages → net output zero. Displacement from null → unbalanced output. Amplitude proportional to displacement, phase indicates direction.
-
Output: AC voltage (needs demodulation for DC output). Infinite resolution (theoretical), no electrical contact.
-
Advantages: Frictionless, long life, high reliability, linear over small range.
-
DiagramSEARCH: "LVDT construction and working principle diagram"
-
-
RVDT (Rotary): Same principle for angular displacement.
-
4.2 Pressure Measurements
-
Fluid Column Manometers (DEC 2024, JUN 2023, NOV 2023):
-
Principle: Hydrostatic balance. $$\displaystyle P = \rho g h $$ (for simple U-tube). Differential pressure $$\displaystyle \Delta P = (\rho_m - \rho_f)gh $$.
-
Types:
-
U-tube: Simple, measures gauge/differential pressure.
-
Inclined: Amplifies reading (longer scale), for low pressures.
-
Well-type: One leg large well, for high pressures.
-
Differential: Two pressures applied to two legs.
-
-
DiagramSEARCH: "U-tube manometer inclined well-type differential manometer diagrams"
-
-
Bourdon Tube (C-type, Helical, Spiral):
-
Principle: Flattened tube tends to straighten under pressure. Motion converted to pointer via gear.
-
Construction: Curved tube, closed end (pressure port), free end (mechanical linkage).
-
Applications: Wide range pressure gauges.
-
-
Bellows Gauge (JUN 2023):
-
Principle: Pressure causes elastic bellows to expand. Expansion moves pointer.
-
Construction: Corrugated bellows, mechanical linkage.
-
Applications: Medium pressure, low pressure, differential pressure.
-
-
Diaphragm Gauge (NOV 2023):
-
Principle: Pressure deflects diaphragm (flat or corrugated). Deflection measured capacitively, piezoresistively, or mechanically.
-
Applications: Low pressure, differential pressure, vacuum.
-
-
Piezoelectric Transducer (JUN 2022, JUN 2023):
-
Principle: Certain crystals (Quartz, PZT) generate surface charge when strained (direct piezoelectric effect). $$\displaystyle Q = d \cdot F $$, $$\displaystyle V = g \cdot t \cdot P $$.
-
Characteristics: High frequency response, AC output only (charge leaks away), very high output impedance (needs charge amplifier). For dynamic pressure only.
-
DiagramSEARCH: "piezoelectric pressure transducer diagram"
-
-
Dead Weight Tester (JUN 2025): (See 1.3)
4.3 Temperature Measurements
-
Thermometric Expansion:
-
Glass Thermometer (JUN 2023): Liquid (Hg, alcohol) in calibrated glass capillary. Simple, direct reading. Limitations: Fragile, limited range (Hg: -38 to 350°C), no remote output.
-
Bimetallic Thermometer (DEC 2024, JUN 2023, JUN 2024):
-
Principle: Two metals with different $\alpha$ bonded. Temperature change causes bending.
-
Construction: Strip (helix/spiral for amplification). Bending moves pointer.
-
Curvature Formula (NOV 2023): $$\displaystyle \frac{1}{r} = \frac{6(\alpha_1 - \alpha_2)\Delta T}{t(1+m)^2} $$ (approx), where $t$=total thickness, $m$=modulus ratio.
-
Applications: Industrial thermostats, dial thermometers.
-
-
-
Electrical Resistance Thermometry:
-
RTD (JUN 2024, JUN 2022): $$\displaystyle R_T = R_0 [1 + \alpha T + \beta T^2 + ...] $$. PT100 ($$\displaystyle R_0=100\Omega $$ at 0°C). Materials: Platinum (best stability/linearity), Nickel, Copper.
-
Advantages: High accuracy, stability, wide range (-200 to 850°C).
-
Limitations: Self-heating, requires excitation current, lead wire resistance error (use 3/4-wire compensation).
-
-
Thermistor (DEC 2024 x2, JUN 2023, JUN 2022): $$\displaystyle R_T = R_0 e^{\beta (\frac{1}{T} - \frac{1}{T_0})} $$ (NTC). High sensitivity (large $$\displaystyle \frac{dR}{dT} $$), non-linear, limited range (-50 to 300°C). PTC used as inrush limiter/sensor.
- Comparison (JUN 2022): RTD = linear, stable, wide range, expensive. Thermistor = non-linear, high sensitivity, cheap, limited range.
-
-
Thermoelectric Thermometry (Thermocouple - JUN 2023):
-
Principle: Seebeck Effect. Two dissimilar metals joined at two junctions generate EMF proportional to temperature difference.
-
Key Points: Requires reference (cold) junction at known temperature (compensation needed). Types: J (Fe-CuNi), K (NiCr-NiAl - most common), T (Cu-CuNi), E (NiCr-CuNi), S/R/B (Pt-Rh/Pt - high temp).
-
Advantages: Wide range (-200 to 2300°C), rugged, no excitation.
-
Limitations: Low output (mV), requires cold-junction compensation, susceptible to noise.
-
-
Radiation Pyrometry (JUN 2022):
-
Optical Pyrometer: Compares brightness of hot object to calibrated filament. Disappearing filament type. For high temperatures (>700°C). No contact.
-
Infrared Pyrometer: Measures IR radiation energy. Wider range, faster.
-
4.4 Strain, Load & Force Measurements
-
Resistance Strain Gauge (Core Topic):
-
Principle: Piezoresistive Effect. Strain $\epsilon$ changes resistance $R$.
-
Gauge Factor (GF): $$\displaystyle GF = \frac{\Delta R / R}{\epsilon} $$. For metallic foil: $ GF \approx 2 $.
-
Derivation (JUN 2023, JUN 2022): $$\displaystyle R = \frac{\rho l}{A} $$. $$\displaystyle \frac{\Delta R}{R} = \frac{\Delta \rho}{\rho} + \frac{\Delta l}{l} - \frac{\Delta A}{A} $$. For Poisson's ratio $\nu$, $$\displaystyle \frac{\Delta A}{A} = -2\nu \epsilon $$. $$\displaystyle \frac{\Delta \rho}{\rho} \propto \epsilon $$. Hence $$\displaystyle GF = 1 + 2\nu + \frac{\frac{\Delta \rho}{\rho}}{\epsilon} \approx 2 $$ for metals.
-
Construction: Metallic foil (most common), wire-wound, semiconductor (high GF, high temp. sensitivity).
-
Mounting (DEC 2024): Surface prep (smooth, clean), adhesive bonding (epoxy), paper backing for handling, orientation along principal strain axis.
-
Wheatstone Bridge (JUN 2024): Converts small $\Delta R$ to measurable voltage $$\displaystyle V_o $$.
-
Quarter Bridge (1 active gauge): $$\displaystyle V_o \approx \frac{V_{ex}}{4} \cdot \frac{\Delta R}{R} $$
-
Half Bridge (2 active, opposite arms): $$\displaystyle V_o \approx \frac{V_{ex}}{2} \cdot \frac{\Delta R}{R} $$
-
Full Bridge (4 active): $$\displaystyle V_o \approx V_{ex} \cdot \frac{\Delta R}{R} $$ (max sensitivity, temperature compensation if arranged properly).
-
Temperature Compensation: Use dummy gauge (identical, unstrained) in adjacent arm.
-
-
DiagramSEARCH: "Wheatstone bridge strain gauge quarter half full bridge configurations"
-
-
Load Cell (NOV 2023): Force sensor using strain gauges in a bridge. Types: Compression/tension, beam, canister. Applications: weighing scales, industrial process.
-
Piezoelectric Force Sensor: For dynamic force only (like pressure).
4.5 Velocity & Rotational Speed Measurements
-
Stroboscope (JUN 2025, NOV 2023):
- Principle: Freezes/reverses apparent motion when flash rate $$\displaystyle f_s $$ equals object speed $n$ (rpm) or sub-multiple.
$$ n = 60 \times f_s \quad (\text{if } f_s = n) \quad \text{or} \quad n = \frac{60 \times f_s}{k} \quad (\text{if } f_s = k \cdot n) $$
* **Operation:** Adjust flash rate until object appears stationary/slow.
* **Applications:** Measure rotational speed, observe vibrating/rotating machinery.
-
Electromagnetic Tachometers:
-
DC Tachogenerator: Voltage $V \propto \omega$. Requires contact/rotating contact.
-
AC Tachogenerator (Induction): Output frequency $f \propto \omega$. No brushes.
-
Proximity/Eddy Current: Non-contact. Measures speed via passing ferromagnetic target.
-
-
Mechanical Tachometers (JUN 2022): Centrifugal (flyball), vibrating reed. Disadvantages: Wear, friction, low accuracy, contact needed.
5.0 CONTROL SYSTEMS FUNDAMENTALS
5.1 Basic Definitions
-
Control System: Interconnected components to achieve desired response.
-
Plant: The part being controlled (e.g., boiler, motor).
-
Input (Command/Reference): Desired value ($r(t)$).
-
Output (Controlled Variable): Actual value ($c(t)$).
-
Disturbance: Unwanted input affecting output ($d(t)$).
-
Manipulated Variable: Input to plant adjusted by controller ($u(t)$).
-
Controller: Compares input & output, generates control action.
5.2 Open-Loop vs. Closed-Loop (Feedback)
-
Open-Loop (JUN 2025, DEC 2024, JUN 2023, JUN 2022, NOV 2023):
-
Block Diagram: Input → Controller → Plant → Output. No feedback path.
-
Operation: Output does not influence control action.
-
Examples: Washing machine timer, traffic light (fixed timing), toaster.
-
Advantages: Simple, stable, low cost.
-
Limitations: No correction for disturbances, sensitive to parameter changes (calibration critical).
-
-
Closed-Loop (Feedback - JUN 2025, NOV 2023):
-
Block Diagram (Standard):
r(t) → [Σ] → Controller → Plant → c(t) ↑ ↓ ←──── Sensor ←──Error $$\displaystyle e(t) = r(t) - c(t) $$.
-
Operation: Output measured, compared to input, error drives controller.
-
Examples (DEC 2024, NOV 2023): Boiler water level control (float/level sensor → controller → feed valve). Temperature control (thermostat), position control (servo).
-
Advantages: High accuracy, rejects disturbances, reduces sensitivity to parameter variations.
-
Limitations: More complex, potential instability (oscillations), higher cost, need sensor maintenance.
[!TIP] Exam Focus: Be ready to draw block diagrams for boiler water level control (DEC 2024, NOV 2023) and temperature control.
-
5.3 Block Diagram Representation (JUN 2025, NOV 2023)
-
Basic Elements: Arrow (signal), Summing point (Σ), Take-off point, Block (transfer function $G(s)$).
-
Interconnections:
-
Series: $$\displaystyle G_1 G_2 $$
-
Parallel: $$\displaystyle G_1 + G_2 $$
-
Feedback: Forward path $G(s)$, Feedback path $H(s)$. Closed-loop TF: $$\displaystyle \frac{C(s)}{R(s)} = \frac{G(s)}{1 + G(s)H(s)} $$ (negative feedback).
-
-
Construction: Identify input, output, sensor, controller, plant, disturbance path. Draw blocks and connect with arrows/summing points.
5.4 Control System Classification
-
By Control Action:
-
On-Off (Two-position): Controller output is either max or min (e.g., thermostat). Simple, causes hunting.
-
Proportional (P): $$\displaystyle u(t) = K_P e(t) $$. Steady-state error usually exists.
-
Integral (I): $$\displaystyle u(t) = K_I \int e(t) dt $$. Eliminates steady-state error.
-
Derivative (D): $$\displaystyle u(t) = K_D \frac{de(t)}{dt} $$. Anticipates error, improves stability.
-
PID: Combination $$\displaystyle u(t) = K_P e + K_I \int e dt + K_D \frac{de}{dt} $$.
-
-
By I/O: SISO (Single-Input Single-Output), MIMO (Multi-Input Multi-Output).
-
By Nature: Linear/Non-linear, Time-invariant/Time-varying, Continuous/Discrete (sampled).
5.5 Specific Control System Examples
-
Servomechanism (JUN 2022): Feedback system where output is mechanical position/velocity. Uses sensors (pot, encoder), amplifier, motor. Example: antenna positioning, CNC machine axis.
-
Temperature Control System (JUN 2022): Heater (plant), thermocouple/RTD (sensor), controller (PID), setpoint. May have disturbance (ambient change, load change).
-
Position Control System (JUN 2022): Similar to servomechanism. Motor drives load, position sensor (LVDT, encoder) provides feedback.
-
Boiler Water Level Control (DEC 2024, NOV 2023):
-
Objective: Maintain constant water level in boiler drum.
-
Elements: Level sensor (float, differential pressure), controller (PID), manipulated variable = feedwater valve opening.
-
Disturbances: Steam demand changes, feedwater pressure changes.
-
DiagramSEARCH: "boiler drum water level control system block diagram"
-
6.0 MATHEMATICAL MODELING OF SYSTEMS
6.1 General Approach
Apply physical laws (Newton's, Kirchhoff's) to system boundaries. Define inputs, outputs, state variables. Assume linearity & time-invariance for transfer function.
6.2 Modeling of Electrical Systems (JUN 2024)
-
Component Equations:
-
Resistor: $$\displaystyle v_R = i R $$
-
Capacitor: $$\displaystyle i_C = C \frac{dv_C}{dt} $$ or $$\displaystyle v_C = \frac{1}{C} \int i_C dt $$
-
Inductor: $$\displaystyle v_L = L \frac{di_L}{dt} $$ or $$\displaystyle i_L = \frac{1}{L} \int v_L dt $$
-
-
Network Analysis: Apply KVL (sum voltages = 0) and KCL (sum currents = 0) to write differential equations.
-
Transfer Function (TF): $$\displaystyle G(s) = \frac{L\{output(t)\}}{L\{input(t)\}} $$ assuming zero initial conditions. TF is ratio of polynomials in Laplace variable $s$.
-
Analogies:
-
Force-Voltage (Mobility) Analogy: Force ↔ Voltage, Velocity ↔ Current, Mass ↔ Capacitor, Spring ↔ Inductor, Damper ↔ Resistor.
-
Force-Current (Impedance) Analogy: Force ↔ Current, Velocity ↔ Voltage, Mass ↔ Inductor, Spring ↔ Capacitor, Damper ↔ Resistor.
-
6.3 Modeling of Mechanical Systems (JUN 2024)
-
Translational:
-
Mass ($m$): $$\displaystyle F = m \frac{d^2x}{dt^2} $$
-
Spring ($k$): $$\displaystyle F = k x $$
-
Damper ($b$): $$\displaystyle F = b \frac{dx}{dt} $$
-
-
Rotational:
-
Inertia ($J$): $$\displaystyle T = J \frac{d^2\theta}{dt^2} $$
-
Torsional spring ($$\displaystyle k_t $$): $$\displaystyle T = k_t \theta $$
-
Rotational damper ($$\displaystyle b_r $$): $$\displaystyle T = b_r \frac{d\theta}{dt} $$
-
-
Write equations using Newton's 2nd law (sum forces/torques = mass/inertia × acceleration).
6.4 Transfer Function Examples
-
First-Order (RC circuit): $$\displaystyle V_o(s)/V_i(s) = \frac{1}{RCs + 1} $$, $$\displaystyle \tau = RC $$.
-
Second-Order (Spring-mass-damper): $$\displaystyle X(s)/F(s) = \frac{1}{ms^2 + bs + k} $$. Standard form: $$\displaystyle \frac{\omega_n^2}{s^2 + 2\zeta\omega_n s + \omega_n^2} $$.
7.0 SPECIALIZED & SHORT NOTE TOPICS
-
Flow Measurement:
-
Venturimeter (DEC 2024, NOV 2023): Converging section, throat, diverging section. Based on Bernoulli & continuity. $$\displaystyle Q = C_d A_t \sqrt{\frac{2(P_1 - P_2)}{\rho(1 - (A_t/A_1)^2)}} $$. Low permanent pressure loss, expensive.
-
Orifice Meter (DEC 2024, NOV 2023): Simple orifice plate. Higher pressure loss, prone to wear, cheaper. Same principle as venturi.
-
DiagramSEARCH: "venturimeter and orifice meter flow measurement diagram"
-
-
Bimetallic Thermometer (DEC 2024, JUN 2023): (See 4.3.1)
-
Stroboscope (JUN 2025, NOV 2023): (See 4.5.1)
-
Load Cell (NOV 2023): (See 4.4.2)
-
Manometer (NOV 2023): (See 4.2.1)
-
Ionisation Transducer (NOV 2023): Measures pressure in vacuum systems. Gas ionized by current, ion current ∝ pressure.
-
Photo-electric Transducer (JUN 2022): Light intensity → electrical signal (photocell, photodiode, phototransistor).
-
Sling Psychrometer (JUN 2022): Measures dry & wet bulb temperatures. Psychrometric chart gives relative humidity from $$\displaystyle T_{dry} $$, $$\displaystyle T_{wet} $$, and atmospheric pressure.
-
Accelerometer Calibration (JUN 2022): Shake on vibration table of known frequency/amplitude, compare output.
-
Impulse Function (JUN 2023): $\delta(t)$: Infinite amplitude, zero width, area=1. $$\displaystyle \int_{-\infty}^{\infty} \delta(t) dt = 1 $$. $$\displaystyle f(t)\delta(t-a) = f(a)\delta(t-a) $$. Used for system identification (output = TF * input).
[!TIP] Final Exam Strategy: For "short note" questions, structure as: 1. Principle/Definition, 2. Construction/Key Components, 3. Working/Operation, 4. Applications, 5. Advantages/Limitations. Use diagrams where possible. For calculation problems (error, GF, bimetallic), show formula, substitute, box answer.